Instruments

Available Instrumentation

Due to time-intensive nature of the laboratory investigations and the large enrollment, not every student will be participating in every listed lab. Instead, all students will conduct lab experiments focusing on DSP with coaxial HPGe detectors, then will be responsible for selecting from the list of potential experiments for the third lab report.

In addition, the final project for the course requires students to propose an instrumentation-focused study. Below is an (incomplete) list of instruments and measurement systems that are available for NE 204. Available relevant information is included for some systems - updates will continue as more information becomes available. Some of the instruments are a good fit for one or more of the experiment writeups, while others are not (yet) explicitly tied to any particular experiment. Please consider these instruments for the rotating labs and final project proposals.

LaBr

LaBr
Description: 2 in. x 2 in. LaBr scintillator with 3 in. Hamamatsu r6233-100
- Operating Voltage: +540V
- Ener. Res. @ 662 keV: 2.6% (from manufacturer)
Location: Etcheverry Hall (1110B or 1140)
Readout: Analog PMT base + SIS | Analog PMT base + NIM | Digibase | SIS33XX
Related Experiments: Lab 3, Time resolution (cf. NE104 Lab 6)
Status: Working as of 10/2/2018
Additional Documentation: None

CCI-2 (VCI)

CCI2
Description: 2x Double-sided strip HPGe detector
- 2 planar detectors, each 15.1 mm thick, segmented orthogonal-strip electrodes
- 152 Strips, 2mm pitch, 0.25mm gap, 5mm guard ring
- Operating Voltage: +800V
- Ener. Res. @ 662 keV: 0.19 - 0.5% (depending on strip)
Location: 1140 Etcheverry Hall
Readout: SIS3302/3150
Related Experiments: Lab 5, Lab 6, Lab 8
Status: Hopefully available by 10/18/18
Additional Documentation: CCI2 TechReport 1, CCI2 TechReport 2

UCBGe2_2008 (re-fabbed in 2011)

Description: Double-sided strip HPGe detector
- 11 mm thick, segmented orthogonal-strip electrodes
- 76 Strips (38 per side), 2mm pitch, 0.25mm gap, 5mm guard ring
- Operating Voltage: +1000V
- Ener. Res. @ 662 keV: 0.19 - 0.5% (depending on strip)
Location: LBNL SDL (70A)
Readout: SIS3302/3150
Related Experiments: Lab 5, Lab 6, Lab 8
Status: Unavailable for Fall 2018
Additional Documentation: UCB2008 Ge2 Refabrication Report (2011)

UCBGe3_2011

Description: Double-sided strip HPGe detector
- 15.1 mm thick, segmented orthogonal-strip electrodes
- 152 Strips, 2mm pitch, 0.25mm gap, 5mm guard ring
- Operating Voltage: See docs
- Ener. Res. @ 662 keV: See docs
Location: LBNL SDL (70A) Readout: SIS3302/3150
Related Experiments: Lab 5, Lab 6
Status: Only half instrumented
Additional Documentation: None

Large Area Imager

LAI
Description: 10x10 NaI(Tl) Coded Aperture Imager
- 10x10 array of NaI(Tl) detectors w/ PMTs
- 4x4x4 in. NaI(Tl) crystals
- Coded aperture array, random configuration, 1 in. thick Pb elements
- Operating Voltage: <1200V (varies by channel)
- Ener. Res. @ 662 keV: ~9% (varies by channel)
Location: 1140 Etcheverry Hall
Readout: SIS3302/3150
Related Experiments: Lab 8 (CA)
Status: Available, 96/100 channels working
Additional Documentation: None

Liquid Scintillation Detector (LSD) Scatter Camera

Description: 2-plane, 24 element kinematic neutron imager - 24 detectors (12 per plane)
- 2 in. x 2 in. EJ309 (xylene-based) liquid scintillators
- Operating Voltage: ~ -1350V (varies by channel)
- Ener. Res. @ 662 keV: N/A
Location: 1140 Etcheverry Hall, 1140B
Readout: SIS3320 | SIS3316
Related Experiments: Lab 4 Lab 9
Status: 3320 Readout available, 3316 Readout requires setup
Additional Documentation: None

Pixelated CZT

CZT
Description: 10 mm3 CZT w/ 3x3 pixellated anode
- 1 cm3 crystal, 3x3 pixellated anode + guard ring (see docs)
- Operating Voltage: ~ -1000V (see docs)
- Ener. Res. @ 662 keV: see docs
Location: 1140 Etcheverry Hall, modular room
Readout: SIS3302/3150
Related Experiments: Lab 7
Status: 2 Detectors Available
Additional Documentation: UCBCZT1, UCBCZT2

PSD Plastics

Description: 1 in. or 2 in. Plastic Scint. w/ PSD
- PSD-capable plastic scintillators from Dr. Natalia Zaitseva, LLNL
Location: 1140 Etcheverry Hall, modular room
Readout: SIS3302/3150
Related Experiments: Lab 4 (PSD)
Status: Un-instrumented, PMTs/bases on order
Additional Documentation: None

CLYC Detector

CLYC
Description: 2x2 in. CLYC Detector
- PSD-capable detector for n/gamma discrimination
- Moderate (~4%) Energy resolution
Location: 1140 Etcheverry Hall
Readout: Bridgeport Systems eMorpho
Related Experiments: Lab 4 (PSD)
Status: Integrated eMorpho PMT/Base Operational w/ working software
Additional Documentation: CLYC

LSO Array & Coded Aperture

LSO
Description: 16 LSO Modules, 4 PMTs per module
- Subset of array formerly used for a PET system
- Excellent time response (O(100 ps))
- Includes 2D coded-aperture mask for high-energy gamma-ray imaging
Location: 1140 Etcheverry Hall, modular room
Readout: SIS3316 (Ethernet or SIS3150 readout)
Related Experiments: Lab 3 (Timing), Lab 8 (CA)
Status: 3150 for off-line readout, 3316 real-time readout under development
Additional Documentation: None

Plastic/Lead System